VEGABAR 64 Pressure transmitter with front-flush CERTEC® For level and pressure measurements in paper, chemical and pharmaceutical as well as water/waste water industries - Dry, ceramic-capacitive sensor element - High abrasion and overload resistance - Deviation in characteristics 0.075% - Adjustment with PLICSCOM or PC - Free-of-charge customer-specific adjustment - Instrument from the plics® family Approval without Ship approval ATEX II 1G, 1/2G, 2G EEx ia IIC T6 ATEX II 1/2D, 2D IP6X T Process fitting / Material Thread G1½A PN60 / 316L Tri-Clamp 2" PN10 / 316L Hyg. connection w. compression nut F40 PN40 / 316L Tuchenhagen Varivent DN32..,1½..,PN25 / 316L Bolting DN40PN40 DIN11851 / 316L SMS DN38 PN6 / 316L DRD PN40 / 316L M44x1,25,Alu-pressure screw PN25 / 316L M44x1,25,pressure screw PN60 / 316L Flange DN40PN40 Form C,DIN2501 / 316L Flange 2"150lb RF,ANSI B16.5 / 316L Seal mesuring cell FKM (Viton) Kalrez 6375 EPDM Chemraz 535 Pressure / Measuring range rel. / 0...2,5 bar (0...250 kPa) rel. / 0...60 bar (0...6000 kPa) rel. / -1...10 bar (-100...1000 kPa) rel. / -1...60 bar (-100...6000 kPa) abs. / 0...5 bar (0...500 kPa) Electronics 4...20mA/HART® Profibus PA Foundation Fieldbus Housing / Protection Plastic / IP66/IP67 Aluminium / IP66/IP67 Aluminium double chamber / IP66/IP67 StSt (electropolished) 316L /IP66/IP67 PE-cable axial IP68, ext. housing plastic IP65 Cable entry / Plug connection M20x1,5 / without ½NPT / without Indicating/adjustment module (PLICSCOM) without mounted